GPIB Operation Description (Option)
: SOURce: SAFEty: STEP: AC: LIMit: REAL: HIGH
This command sets selected STEP which AC withstand voltage real
current high limit value.
Example : Input command “SAFE:STEP 1:AC:LIM:REAL 0.0001”
Description: This command means AC withstand voltage real current high
limit value of the main unit STEP1 is 0.10mA.
[: HIGH]?
: SOURce: SAFEty: STEP: AC: LIMit: REAL: HIGH?
This command queries selected STEP which AC withstand voltage real
current high limit value.
Example : Input command “SAFE:STEP 1:AC:LIM:REAL?”
Return message “1.000000E-04”
Description: Return message “1.000000E-04” means ARC checking
value of the main unit STEP 1 is set to 0.10mA.
: TIME
: RAMP <numeric_
value>
: SOURce: SAFEty: STEP: AC: TIME: RAMP
This command sets selected STEP which test ramps to setting voltage
needed time.
Example : Input command “SAFE:STEP 1:AC:TIME:RAMP 5”
Description: This command means test ramps to setting voltage needed
time of the main unit STEP 1 is 5.0sec.
: RAMP?
: SOURce: SAFEty: STEP: AC: TIME: RAMP?
This command queries selected STEP which test ramps to setting voltage
needed time.
Example : Input command “SAFE:STEP 1:AC:TIME:RAMP?”
Return message “5.000000E+00”
Description: Return message “5.000000E+00” means test ramps to
setting voltage needed time of the main unit STEP 1 is set to
5.0sec.
[: TEST] <numeric_
value>
: SOURce: SAFEty: STEP: AC: TIME: TEST
This command sets selected STEP which test needed time.
Example : Input command “SAFE:STEP 1:AC:TIME 10”
Description: This command means test needed time of the main unit
STEP 1 is 10.0sec.
[: TEST]?
: SOURce: SAFEty: STEP: AC: TIME: TEST?
This command queries selected STEP which test needed time.
Example : Input command “SAFE:STEP 1:AC:TIME?”
Return message “1.000000E+01”
Description: Return message “1.000000E+01” means test needed time of
the main unit STEP 1 is set to 5sec.
: FALL <numeric
value>
: SOURce: SAFEty: STEP: AC: TIME: FALL
This command sets selected STEP which setting voltage value fall to
zero needed time.
5-21