HIPOT Tester 19051/19052/19053/19054 User’s Manual
Example : Input command “SAFE:STEP 1:AC:TIME:FALL 4”
Description: This command means setting voltage value of the main unit
STEP 1 fall to zero needed time is 5.0 sec.
: FALL?
: SOURce: SAFEty: STEP: AC: TIME: FALL?
This command queries selected STEP which setting voltage value fall to
zero needed time.
Example : Input command “SAFE:STEP 1:AC:TIME:FALL?”
Return message “4.000000E+00”
Description : Return message “4.000000E+00” means setting voltage
value of the main unit STEP 1 fall to zero needed time is
4 sec.
: CHANnel
[: HIGH] <channel_list>
: SOURce: SAFEty: STEP: AC: CHANnel: HIGH
This command sets output channel status of scanning test high voltage.
Example : Input command “SAFE:STEP 1:AC:CHAN(@(1,3))”
Description: This command means output channel of scanning test of
the main unit STEP 1 is set to channel 1 and 3 HIGH output.
Example : Input command “SAFE:STEP 1:AC:CHAN(@(0))”
Description: This command means HIGH output channel of scanning test
output channel of the main unit STEP 1 is set to OFF.
[: HIGH]?
: SOURce: SAFEty: STEP: AC: CHANnel: HIGH?
This command queries output channel status of scanning test high
voltage.
Example : Input command “SAFE:STEP 1:AC:CHAN?”
Return message “(@(1,3))”
Description: Return message “(@(1,3))” means output channel status of
scanning test of the main unit STEP 1 is channel 1 and 3
HIGH output.
: LOW <channel_list>
: SOURce: SAFEty: STEP: AC: CHANnel: LOW
This command sets output status of scanning common test channel
(RTN/LOW).
Example : Input command “SAFE:STEP 1:AC:CHAN:LOW (@(2,4))”
Description: This command means output channel of scanning test of the
main unit STEP 1 is set to channel 2 and 4 LOW output.
Example : Input command “SAFE:STEP 3:AC:CHAN:LOW (@(0))”
Description: This command means LOW output channel of scanning test
of the main unit STEP 1 is set to OFF.
: LOW?
: SOURce: SAFEty: STEP: AC: CHANnel: LOW?
This command queries output status of scanning common test channel
(RTN/LOW).
Example : Input command “SAFE:STEP 1:AC:CHAN:LOW?”
Return message “(@(2,4))”
Description: Return message “(@(2,4))” means output channel status of
5-22