GPIB Operation Description (Option)
scanning test of the main unit STEP 1 is channel 2 and 4
LOW output.
: DC
[: LEVel] <numeric_value>
: SOURce: SAFEty: STEP: DC: LEVel
This command sets selected STEP which DC withstand voltage test
needed voltage value.
<numeric_value> is 50-6000 (included), the unit is volt.
Example : Input command “SAFE:STEP 2:DC 4000”
Description : This command means DC withstand voltage test needed
voltage value of the main unit STEP 2 is set to 4000V.
[: LEVel]?
: SOURce: SAFEty: STEP: DC: LEVel?
This command queries selected STEP which DC withstand voltage test
needed voltage value.
Return value is 50-6000(included), the unit is volt.
Example : Input command “SAFE:STEP 2:DC?”
Return message “4.000000E+03”
Description: Return message “4.000000E+03” means DC withstand
voltage test setting voltage value of the main unit STEP 2 is
4000V.
: LIMit
[: HIGH] <numeric_value>
: SOURce: SAFEty: STEP: DC: LIMit: HIGH
This command sets selected STEP which DC withstand voltage leakage
current high limit.
Example : Input command “SAFE:STEP 2:DC:LIM 0.002999”
Description: This command means DC withstand voltage leakage current
high limit of the main unit STEP 2 is set to 2.999mA.
[: HIGH]?
: SOURce: SAFEty: STEP: DC: LIMit: HIGH?
This command queries selected STEP which DC withstand voltage
leakage current high limit value.
Example : Input command “SAFE:STEP 2:DC:LIM?”
Return message “2.999000E-03”
Description: Return message “2.999000E-03” means DC withstand
voltage leakage current high limit value of the main unit
STEP 2 is 2.999mA.
: LOW <numeric_value>
: SOURce: SAFEty: STEP: DC: LIMit: LOW
This command sets selected STEP which DC withstand voltage leakage
current low limit.
Example : Input command “SAFE:STEP 2:DC:LIM:LOW 0.000001”
Description: This command means DC withstand voltage leakage current
low limit of the main unit STEP 2 is set to 0.001mA.
: LOW?
: SOURce: SAFEty: STEP: DC: LIMit: LOW?
This command queries selected STEP which DC withstand voltage
5-23