Dew point sensor FA5xx
LD500 V2.03 Seite 44 von 67
13.5.1.3 Definition of Reference pressure (absolute pressure value)
External sensor ➔ Settings ➔ Sensor settings ➔ C1➔ arrow right (2.page)➔Pressure Setting ➔
Text field Ref.Pressure
13.5.1.4 Field adjustment
External sensor ➔ Settings ➔ Sensor settings ➔ C1➔ arrow right (2.page)➔ Field adjustment
Here, a one-point adjustment can be
performed.
For that purpose, please enter in the text
box "Reference Value" the new correct dew
point value.
Then by pressing the "Adjustment" button
taking over the inserted reference value.
Calibration can be put back to factory
setting by pressing „Reset“.
For each performed calibration, the
counter is increased by 1.
Reference pressure is the pressure for that
the dew point in relaxation will be back-
calculated.
Default- Value is 1013 mbar (Atm.
Pressure).
Confirm the settings by pressing the OK
button.