MAINTENANCE
306
AV500/AV900 2D CAMERA
Tighten Mounting Hardware
1. Check all cameras, Deflection Mirrors, CBX connection boxes, and power supply
mounting hardware as applicable. Tighten as necessary. Do not over-tighten. Be
sure not to disturb the equipment’s alignment as it relates to the product trans-
port.
2. Check the mounting hardware of the Photoelectric Sensor (if
this option is being
used). Tighten as necessary. Do not over-tighten.
3. Check the mounting hardware of the tachometer (if this option is being used).
Tighten
as necessary. Do not over-tighten.
Checking Camera System Connections
This is a safety check recommended for harsh environments where vibration may be a
problem. (See “Electrical Installation” on page 46)
1. Check all wiring connections to the camera and illumination connector panels.
Tight
en any loose connections as necessary. Do not over-tighten.
2. Check all wiring connections to the CBX connection box. Tighten any loose connec-
tions as necessary. Do not over-tighten.
3. Check all cabling/conduit for signs of wear/damage. Repair/replace any damaged
c
abl
e connections as necessary.
Verify Camera Operation
If after performing maintenance, the imaging system continues to perform below the
normal operations experienced with the system under normal daily conditions, contact
Datalogic through our website at www.datalogic.com.
Verify Photoelectric Sensor Operation
1. Block the Photoelectric Sensor emitter beam to confirm it is aligned properly with
the reflector.
2. Verify that the TRIGGER LED on camera lights
when photoelectric sensor beam is
blocked.
3. If the photoelectric sensor’s LED does not change status, adjust the photoelectric
sensor so tha
t it is properly aligned with the reflector.
4. If the TRIGGER LED on the barcode reader does not light when the photoelectric
sensor’s beam i
s blocked, check the cabling between the photoelectric sensor and
CBX connection box for damage.