EasyManuals Logo

DATREND Systems vPad-A1 User Manual

DATREND Systems vPad-A1
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
vPad-A1 OPERATORS MANUAL
• Tall T-Wave Test 0 to ± 150% in steps of 10% of Amplitude Setting
(0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.1, 1.2, 1.3,
1.4 and 1.5 mV at 1 mV)
• ST Segments: ± 0, 5, 10, 20, 30, 40, 50, 60, 70, 80 % of Amplitude
Setting
(0, -0.05, -0.1, -0.2, -0.3, -0.4, -0.5, -0.6, -0.7, -0.8,
0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7 and 0.8 mV at 1 mV)
• Performance
Amplitude:
0.05 to 0.5 mV in 0.05 mV steps, 0.5 to 5.0mV in 0.25 mV
steps
(applies to all performance waveforms)
Respiration:
• Rates: 15 to 150 BrPM incremental with 15 customizable
presets
• Baseline Impedance: 500, 1000, 1500, 2000 ohms on Leads I, II, III
• Apnea Selections: 12, 22, 32 seconds, and continuous (0 BPM)
• Impedance Variation: 0 to 1.0 ohms in 0.05 ohm steps, 1.0 to 5.0 in 0.25 ohm
steps
• Simulation Types: Inspiration:Expiration ratios of 1:5, 1:4, 1:3, 1:2, and
1:1, and Ventilated 1:1
• Lead Reference: LA or LL
Temperature:
• Channels: 2
• Body Temperatures: 20 to 42°C in 0.5°C steps
• Probe Compatibility: YSI series 400 or 700
• Accuracy: ± 0.03°C , ± 0.01°C at 30, 32, 35, 37, 40, 42°C
Pacemaker:
Specifications/Chapter 1 # Page 6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DATREND Systems vPad-A1 and is the answer not in the manual?

DATREND Systems vPad-A1 Specifications

General IconGeneral
BrandDATREND Systems
ModelvPad-A1
CategorySwitch
LanguageEnglish