vPad-A1 OPERATORS MANUAL
When a valid AutoSequence has been selected, the
button will launch the Equipment Information form. It allows users
to enter details on the equipment being tested, as described in
section 5.1.2.3 -
Entering Equipment Information
. To delay this
step until the end of the AutoSequence, press
instead.
After the
Equipment Information form is dismissed, the A1 begins
iterating through the Groups that comprise the AutoSequence.
The test will sequentially activate each of the AutoSettings in the
first Group, then go back and display the Test Result screen for
each simulation or performance test setup in the same
chronological order.
For instance, the sample
AutoSequence “Hypertensive” has one
Group of AutoSettings, which contains setup information for ECG,
Respiration, Temperature, IBP, SpO2 and NIBP. The ECG
AutoSetting
is first in the Group, so the A1 starts by displaying the
ECG Test Result interface in
Figure 126
.
Figure 125
Figure 126
Automation and Test Records/Chapter 5 # Page 82