GUIDANCE AND MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The Scanmate Flex is intended for use in the electromagnec environment specied below. The customer
or the user of the Scanmate Flex should assure that it is used in such an environment.
Immunity Test Test Level
Compliance
level
Electromagnec environment - guidance
Conducted RF
Immunity
IEC 61000-4-6
Radiated RF
Immunity
IEC 61000-4-3:2006
A1:2007 A2:2010
6 Vrms
150 kHz to 80 MHz
for AC Mains
6 Vrms
150 kHz to 80 MHz
for signal I/O lines
3 V/m
80 MHz to
1000 MHz
3 V/m
1000 MHz to
2000 MHz
3 V/m
2000 MHz to
2500 MHz
Complies
Complies
Portable and mobile RF communicaons
equipment should be used no closer to any
part of the Scanmate Flex, including cables,
than recommended separaon distance
calculated from the equaon applicable to
the frequency of the transmier.
Recommended separaon distance
150 kHz—80 MHz
80 MHz—800MHz
800 MHz—2.5 GHz
where P is the maximum output power
rang of the transmier in was (W)
according to the transmier manufacturer
and d is the recommended separaon
distance in meters (m).
Field strengths from xed RF transmiers,
as determined by an electromagnec site
survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situaons. Electromagnec propagaon is aected by
absorpon and reecon from structures, objects and people.
a
Field strengths from xed transmiers, such as base staons for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theorecally with accuracy. To assess the electromagnec environment due to xed RF
transmiers, an electromagnec site survey should be considered. If the measured eld strength in
the locaon in which the Scanmate Flex is used exceeds the applicable RF compliance level above,
the Scanmate Flex should be observed to verify normal operaon. If abnormal performance is
observed, addional measures may be necessary, such as re-orienng or relocang the Scanmate
Flex.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than [V
1
] V/m