Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 161
Alarms
inAgent high --- After 2 consecutive
respiratory phases and
15 seconds
or
after 30 seconds if no
respiratory phases are
detected
>165 seconds later
Inspiratory xMAC high --- Insp. MAC ≥3
for more than
180 seconds
longer than 30 seconds:
insp. MAC ≥3
and
exp. MAC ≥2.5
or
insp. MAC ≥5
xMAC low 0 to 60 seconds >60 s ---
Apnea (no CO
2) --- At the latest after
20 seconds
(for RR ≥6)
or
At the latest after
35 seconds
(for RR <6)
or
at the latest after
65 seconds for the
Pause, Manual / Sponta-
neous and Ext. FGO
modes
15 seconds later
(for RR ≥6)
or
30 seconds later
(for RR <6)
No CO
2 detected >60 s --- ---
O
2 measurement not
available
N
2O measurement not
available
Agent measurement
not available
O
2 measurement tem-
porarily inaccurate
CO
2 sensor accuracy
low
Agent measurement
temporarily inaccurate
N
2O measurement
temporarily inaccurate
Measured gas concen-
trations temporarily
inaccurate
>20 s --- ---
Alarm Priority Priority Priority
Note
(Low)
Caution
(Medium)
Warning
(High)