EasyManua.ls Logo

Dräger Atlan A350 - Page 162

Dräger Atlan A350
332 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
162 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Alarms
Gas measurement alarms with inspiratory O2 measurement
Ventilation alarms
Alarm Priority Priority Priority
Note
(Low)
Caution
(Medium)
Warning
(High)
FiO
2 low --- After at least 5 seconds
FiO
2 high --- After at least 5 seconds
O
2 measurement not
available
O
2 measurement tem-
porarily inaccurate
O
2 sensor not ready
>20 s --- ---
Alarm Priority Priority Priority
Note
(Low)
Caution
(Medium)
Warning
(High)
Apnea (no flow)
Apnea (no pressure)
Apnea
--- At the latest after
20 seconds
(for RR 6)
or
At the latest after
35 seconds
(for RR <6)
or
at the latest after
65 seconds for the
Pause, Manual / Sponta-
neous and Ext. FGO
modes
15 seconds later
(for RR 6)
or
30 seconds later
(for RR <6)
Apnea Ventilation At the latest after 20 seconds
(15 seconds for RRmin 4)
(configurable, see page 168)
---
Tidal volume not
achieved
--- After 3 consecutive
breaths
---
Airway press. continu-
ously high
--- --- >15 seconds above the
manually or automati-
cally set limit
Airway pressure nega-
tive
--- --- Pmean < –2
or
Paw < –10
PEEP/CPAP high --- Airway pressure
>(PEEP +5 hPa
(cmH
2O)) during more
than 10 consecutive
breaths
---
Airway pressure not
achieved
--- After 2 consecutive
breaths
---

Table of Contents