Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 173
Configuration
11.3.4.2 Vertical tab "Vent. 2"
11.3.4.3 Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Headline/
Parameter
Setting range
Description
Start settings for ventilation Specifies the start settings for the venti-
lation.
Pmax
[mbar]; [hPa];
[cmH
2O]
12 to 80
40
12 to 80
30
12 to 80
25
Pinsp
[mbar]; [hPa];
[cmH
2O]
7 to 80
15
7 to 80
15
7 to 80
15
∆Psupp
[mbar]; [hPa];
[cmH
2O]
Off;
3 to 80
10
Off;
3 to 80
10
Off;
3 to 80
10
Insp term
[%]
5 to 80
25
5 to 80
25
5 to 80
25
PEEP
[mbar]; [hPa];
[cmH
2O]
Off;
2 to 35
3
Off;
2 to 35
3
Off;
2 to 35
3
Slope
[s]
0 to 2
0.2
0 to 1.5
0.2
0 to 1.5
0.2
RRmin
[/min]
Off;
3 to 25
6
Off;
3 to 25
10
Off;
3 to 25
15
% Tplat
[%]
0 to 60
20
0 to 60
20
0 to 60
20
Trigger
[L/min]
0.3 to 15
4.0
0.3 to 15
2.0
0.3 to 15
1.0
Sync. SIMV; CMV SIMV; CMV SIMV; CMV
Headline/
Parameter
Setting range
Description
Start settings for fresh gas Selects the start settings for the fresh-
gas delivery.
FG O
2
[%]
21 to 100
100
21 to 100
100
21 to 100
100
Sets the O
2 flow.
FG flow
[L/min]
0.20 to
15.00
2.00
0.20 to
15.00
2.00
0.20 to
15.00
2.00
Sets the fresh-gas flow.