176 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.6 System setup > System
11.3.6.1 Vertical tab "General"
Gas supply
Disable N
2OOn; Off A device with connectors for nitrous
oxide can be configured so that nitrous
oxide is no longer displayed, nor can it
be selected as a carrier gas.
Prerequisite:
– The current carrier gas is Air.
– Nitrous oxide is not connected or
available.
Restart the device if necessary to
implement the settings.
On: Nitrous oxide cannot be delivered.
Off: Nitrous oxide can be delivered.
Headline/
Parameter
Setting range
Description
Language List of available languages
English (United States)
Selects the language.
A flag symbol identifies the tabs that
lead to the page with the language set-
tings.
Time source MEDIBUS 1; MEDIBUS 2; NTP server;
None
Selects the source for the time synchro-
nization.
Prerequisite: The connected device
supports this function.
Date and time day; month; year
hour; minute
Sets the date and time. The change is
applied on leaving the "General" verti-
cal tab.
Automatic switch
to daylight sav-
ings time
On; Off Activates or deactivates the automatic
changeover to daylight saving time.
OR working hours Hour : Minute to Hour : Minute
6:30 to 18:30
Sets the working hours of the operating
room.
During this time, the gas measurement
is kept in a pre-warmed and calibrated
state so measured values are available
after only a short waiting period. How-
ever, this decreases the life span of the
patient-gas measurement module.
Device name Device name (up to 16 alphanumeric
characters)
A3XX
Changes the device name in order,
e.g., to enter the operating location.
Headline/
Parameter
Setting range
Description