Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 175
Configuration
11.3.4.5 Vertical tab "General"
11.3.5 System setup > Licenses/Options
11.3.5.1 Vertical tab "Licenses/Options"
Headline/
Parameter
Setting range
Description
Pinsp changes
with PEEP
On; Off Determines whether changing PEEP
automatically changes Pinsp also, so
that the difference between PEEP and
Pinsp remains constant.
Ti changes with
RR (I:E ratio is
locked)
On; Off Determines whether the change in
respiratory rate automatically effects a
change in Ti, so that the I:E.ratio
remains constant.
Breathing system
warmer
On; Off Switches the breathing system warmer
on or off.
Auto Wake-up
(opens start dia-
log upon respira-
tory activity)
On; Off Determines whether the Start dialog
opens automatically when ventilation
activity is detected (e.g., as a result of
repeated squeezing of the breathing
bag).
Anesthetic gas
compensation
Auto; Off;
Des; Sev
Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
2 measurement.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Des 0 to 18
6
Sev 0 to 6.5
2.1
Headline/
Parameter
Setting range
Description
Licenses for software options – Overview of available and active
software options.
– Activating software options.
Observe the information in the following
section: "Activating software
options", page 182