EasyManua.ls Logo

Dräger Atlan A350 - Page 196

Dräger Atlan A350
332 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
196 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
12.6.1 Failure of the integrated patient-gas measurement module
(PGM)
WARNING
Risk due to gas measurement failure
If the gas measurement fails, the patient can no longer be adequately
monitored.
Check the sample line and the water trap for damage or blockages. Pay
attention to the replacement intervals.
Replace the patient-gas measurement module.
Ensure corresponding substitute monitoring.
Arrange for appropriate substitute monitoring conforming to ISO 80601-2-55.
12.6.2 Replacing the patient-gas measurement module (PGM)
Contact service personnel to replace the PGM.
Prerequisite:
There is no patient connected.
The device is switched off.
The device is disconnected from the power supply. To do this, disconnect the
power plug.
The water trap has been removed.
1. WARNING
Incorrect gas measurement
Faulty or non-functional PGMs can cause incorrect gas measurements. As a
result, the patient could be put at risk.
Have the PGM replaced by service personnel.
The O
2 sensor cell is not present or is
not plugged in correctly.
Check the O2 sensor cell:
Insert the sensor cell if it is not pres-
ent.
Check that the sensor cell is cor-
rectly seated.
Calibrate the sensor cell again.
An initialization phase of up to 30
minutes is required after a new O
2
sensor cell is inserted. Only then is
O
2 measurement and calibration
possible once more.
The O
2 sensor cell is faulty or is expir-
ing.
Replace the O2 sensor cell.
Cause Remedy

Table of Contents

Other manuals for Dräger Atlan A350

Related product manuals