Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 197
Troubleshooting
At the left-hand side of the device, take hold of the cover (1) at the top and
bottom and pull it off directly to the front.
2. On the underside, remove the water trap (2) and the RFID antenna, if present.
To do this, gently press the side surfaces (3) of the RFID antenna together. Pull
the RFID antenna off downwards and leave it on the device.
Do not pull on the antenna cable. Do not kink the antenna cable.
3. Remove the 2 screws and the fixing plate (4).
4. Withdraw the patient-gas measurement module from the compartment.
5. Fit a new patient-gas measurement module. Proceed in reverse order to do this.
6. Insert a new water trap.
7. Turn on the device.
8. Perform a software download for the new PGM if required.
9. Perform a system test.