Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 251
Reprocessing
13.11.2.2 Inserting the fresh-gas decoupling valve
The fresh-gas decoupling valve (black) must be fitted on the bottom side of the
lower part of the breathing system.
1. Turn the upper part of the breathing system over.
2. Insert the valve (1) and lock it by turning it clockwise.
13.11.2.3 Fitting the APL valve
1. WARNING
Risk of an incorrectly set pressure limitation
If the lower dot marked on the APL valve is not aligned correctly, the APL valve
may be set or read off incorrectly.
When fitting, take care that the lower dot is facing the user during operation.
Align the APL valve correctly. The lower dot (1) must face the user during
operation.
3709739506
1