252 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
2. Place the APL valve (2) vertically on the upper part of the breathing system (3).
3. Tighten the knurled nut (4).
13.11.2.4 Fitting the upper part of the breathing system
1. Make sure the levers (1) are rotated outward.
2. Place the upper part of the breathing system (2) on the lower part of the
breathing system (3).
3. Turn the levers (1) by approximately 120° clockwise.