Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 277
Technical data
T0...90 <45 s (RR ≥6/min)
<105 s (RR <6 /min)
Leakage minute volume MVleak
Range 0 to 40 L/min
Accuracy ±25 % of measured value or ±0.2 mL/min
(the larger value applies)
Resolution of displayed value 0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV ≥1L/min)
Low-flow wizard
Standardization The following data are standardized to STPD
conditions.
Range 0 to 8 L/min
Accuracy ±25 % of measured value or ±100 mL/min
(the larger value applies)
Respiratory rate
Total RR
Spontaneous RRspon
Mandatory RRmand
Range 0 to 100 /min
Accuracy ±10 % or ±1 /min (the larger value applies)
Resolution of displayed value 1 /min
T
0...90 <45 s (RR ≥6/min)
<105 s (RR <6 /min)
Dynamic compliance
Dynamic compliance Cdyn
Mean dynamic compliance Cdyn mean
Range 0 to 200 mL/hPa (mL/cmH
2O)
Accuracy for gas measurement with PGM ±15 % or ±1 mL/hPa (mL/cmH
2O) (the larger
value applies)
Accuracy without PGM ±40 % or ±3 mL/hPa (mL/cmH
2O) (the larger
value applies)
Information The values may be distorted by spontaneous
breathing, thus reducing the measurement
accuracy.
Resolution of displayed value 0.1 mL/hPa (mL/cmH
2O)
Resistance R
Range 0 to 100 hPa/L/s (cmH
2O/L/s)
Accuracy for gas measurement with PGM ±30 % of the measured value or ±3 hPa/L/s
(cmH
2O/L/s) (the larger value applies)
16.13 Measuring systems and displays (continued)