34 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.2.2 Ventilation drive
The device uses a piston drive as the ventilation drive and is equipped with
compliance compensation and fresh-gas decoupling. For further information see:
"Description of the ventilation drive", page 307.
3.2.3 Gas delivery
The device can deliver mixtures of medical gases to which an anesthetic agent is
added by means of a vaporizer.
Available gas mixtures
–O
2 and Air with 2-gas mixer and 3-gas mixer
–O
2 and N2O with 3-gas mixer
Usable anesthetic agents
Sevoflurane
Desflurane
Isoflurane
Halothane
Enflurane
3.2.4 Ventilation modes
Manual / Spontaneous
CPAP / PSV
–PC - CMV
–PC - SIMV
PC - SIMV / PS
–VC - CMV
–VC - SIMV
VC - SIMV / PS
VC - CMV / AutoFlow
Advanced neonatal sup-
port
Includes advanced ventilation functions and monitor-
ing functions for the ventilation of neonates:
Minimum settable tidal volume of 5 mL
Higher sweep speed
Increased flow measurement sensitivity for more
precise ventilation monitoring
Expert view Provides the following advanced views:
4 waveforms
3 waveforms and one row with parameter fields