Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 41
Overview
Ventilator failure – Manual ventilation or spontaneous breathing possible
– Fresh-gas delivery available
– Anesthetic agent delivery from connected vaporizers
possible
Alternatively:
– Backup manual mode
Screen fault (screen
does not respond to
operation or has
failed)
If backup manual mode is activated:
– Manual ventilation or spontaneous breathing possible
– Emergency O
2 delivery (electronically controlled gas
mixer) or fresh-gas delivery (mechanically controlled
gas mixer) available
– Anesthetic agent delivery from connected vaporizers
possible
Complete device
failure
If backup manual mode is activated:
– Manual ventilation or spontaneous breathing possible
– Emergency O
2 delivery (electronically controlled gas
mixer) or fresh-gas delivery (mechanically controlled
gas mixer) available
– Anesthetic agent delivery from connected vaporizers
possible
Fault Safety function