40 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.2.12.3 Overview
The following table gives an overview of the integrated safety functions which come
into effect if problems occur during operation:
Fault Safety function
Leakage Automatic leakage compensation
The pressure is held constant at the PEEP level.
With pressure-controlled breaths, the pressure is regu-
lated according to the set pressure.
(In all pressure-controlled ventilation modes, small
amounts of leakage are compensated by the piston
drive.)
Mains power supply
failure
Uninterruptible power supply provided by internal battery
Battery operation possible for at least 45 minutes, typi-
cally 150 minutes
Automatic deactivation of the breathing system warmer
to increase battery runtime
Mains power supply
failure and battery
discharged
Manual ventilation and spontaneous breathing are
available
Emergency O
2 delivery (electronically controlled gas
mixer)
Fresh-gas delivery (mechanically controlled gas mixer)
Delivery of anesthetic agents via connected vaporizers
Failure of the central
gas supply system
Use of the connected gas cylinders
Complete failure of
the gas supply
Mechanical ventilation with ambient air possible (the
hose with the breathing bag will have to be removed for
this)
No anesthetic agent delivery from the connected
vaporizers possible; switch to intravenous anesthetic
agent required
Flow measurement
failure
Mechanical ventilation can be continued.
Limitations with regard to displayed measured values,
measurement accuracies, and when triggering manda-
tory breaths are possible.
Failure of fresh-gas
delivery (electroni-
cally controlled
mixer only)
Emergency O
2 delivery
For further information see: "Backup manual mode",
page 38.
Anesthetic agent delivery from connected vaporizers
possible
All ventilation modes are available.
Alternatively:
Backup manual mode

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals