Alarms and Messages
VF4 Infinity Gamma Series Page 5-17
Heart Rate (HR), Arrhythmia (ARR) and ST Segment Analysis (ST)
Displayed Message Possible Cause Suggested Action
Message: Arrhythmia
Relearning
Parameter value: LRN
The monitor is learn-
ing the patient’s nor-
mal QRS complex to
use as reference.
Wait until the message dis-
appears from the screen.
Message: Asystole
Parameter value: ASY
No QRS detection for
the last 4 seconds.
Heart rate is below 15
beats/ minute.
Observe the patient and
treat if clinically indicated.
Select a lead with at least
an amplitude of 0.5 mV.
Select another lead to meet
the above criterion.
Reposition or change the
electrodes if the amplitude
is still low.
Use good skin preparation.
Message: RL Lead Off *
HR parameter value:
<value>
ST<lead1>/ST<lead2>
parameter value: <blank>
Message: V Lead Off *
Message: V+ Lead Off
Message: RA Lead Off
Message: LL Lead Off
Message: LA Lead Off
Message: ECG Leads Off
HR parameter value: ***
ST<lead1>/ST<lead2>
parameter value: <blank>
The monitor has
detected a lead-off
condition for the cur-
rently processed lead.
The cause could be
one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the
electrode
Wrong cable type
selection.
The M
ULTIMED/
NEOMED pod may be
defective.
Check the patient cable and
lead wires carefully.
Check the MULTIMED/
NEOMED pod. Replace pod
if defective.
Replace any cable or lead
wire that is suspect.
Reapply gel on the elec-
trode or change the elec-
trode (if disposable).
Select another ECG lead
for processing if the elec-
trode or lead cannot be
replaced.
* For all the electrode disconnected messages (V, V+ RA, LL, LA, RL) the parameter box will
show *** if this is the lead currently being monitored. If another valid lead is being monitored
(i.e. ECG 1&2) then the parameter value for HR, ST<lead1> and ST<lead2> will be <value>
Message: ARR Cannot
Learn Lead <lead>
Parameter value: <value>
Lead not connected. Check lead connections