Page 5-26 Infinity Gamma Series VF4
Message: etCO
2
Sensor
Temp Not Stable
Parameter value: ***
The sensor temperature is
unstable following warm-
up.
• Wait at least three min-
utes for the message to
disappear. If the mes-
sage persists, replace
the reusable sensor.
Message: Check etCO2
Check Airway Adapter/
Cal.
Parameter value: ***
Airway adapter is dirty, not
fully seated, or out or
calibration.
• Make sure the adapter
is properly seated.
• Clean and calibrate the
airway adapter.
Message: etCO
2
Calibrating Sensor
Parameter value:
<blank>
Calibrating on zero cell. • Informational mes-
sage; no action
required.
Message: etCO
2
Cannot
Cal. Sensor
Parameter value: ***
Calibration on zero cell
could not be completed
because of CAPNOSTAT
temperature instability.
• Check for any heat
sources warming the
sensor and remove
them.
• Wait at least three min-
utes for the tempera-
ture to stabilize.
Message: etCO
2
Adapter
Failure
Parameter value: ***
Airway adapter is dirty, not
fully seated, or out of
calibration.
• Make sure the adapter
is properly seated.
• Clean and calibrate the
reusable airway
adapter.
Message: etCO
2
Place
Sensor On Ref Cell
Parameter value:
<value>
Calibration on zero cell
completed successfully.
• Place sensor on the
reference cell and wait
for calibration to com-
plete.
Message: etCO
2
Sensor
Cal. Failed
Parameter value: ***
Calibration on zero cell
failed.
• Recalibrate. If the mes-
sage persists, try a new
sensor.
Message: etCO
2
Verifying Sensor Cal
Parameter value: ***
Calibrating on reference
cell.
• Informational mes-
sage; no action
required.
Message: etCO
2
Sensor
Cal. Verified
Parameter value:
<value>
Verification completed
successfully.
• Informational mes-
sage; no action
required.
End Tidal CO
2
(etCO
2
), Inspired CO
2
(iCO
2
), and RRc (continued)