Page 5-38 Infinity Gamma Series VF4
Message: NBP Out of Range
Parameter value: ***
NBP values are
above/below the NBP
measuring range.
• Check patient and
treat, if necessary.
• Verify that the inflation
mode is appropriate for
the patient. Change if
necessary.
Message: NBP Low Inflation
Limit
Parameter value: ***
The patient’s systolic
pressure is above the
current inflation limit.
• Change to a higher
inflation limit.
Message: NBP Check Cuff
Size
Parameter value: ***
The wrong size NBP
cuff is being used on
the patient.
• Verify the cuff size and
placement are proper
for the patient.
• Verify all connections.
• Change the cuff if nec-
essary.
Message: NBP Check Hose
Connection
Parameter value: ***
There is no significant
increase in cuff pres-
sure during the infla-
tion cycle.
• Verify all connections.
• Change the hose if
necessary.
Non-Invasive Blood Pressure (NBP) (continued)