Page 14-8 Infinity Gamma Series VF4
NOTES:
! Sampling lines and T-connectors to the patient’s airway are not
reusable. They must be replaced under the following condi-
tions:
-- A new patient is connected to the module.
-- The cleanliness of the tubing/connector is suspect.
! Use only Dräger-approved sidestream sampling lines. Dräger
does not assume responsibility for the reliability and safety of
Scio measurements, if non-approved tubing is used.
WARNINGS:
! Do not use the module without a Watertrap.
! Sampling lines should be kept as short as possible
(but not stretched) to minimize dead space and opti-
mize response time. Long sampling lines degrade
the performance of side stream measurements, may
affect accuracy, and result in slower response times.
! Always use Dräger-approved Scio sampling lines
(polypropylene). Never use standard pressure-sen-
sor tubing (PVC). PVC tubing absorbs anesthetic
agents, which it later releases (degassing). The use
of standard PVC tubing can result in erroneous agent
concentration readings.
! Sampling lines and T-connectors are not reusable
and must be replaced after each patient. Water traps
and fan filters must be replaced at regular intervals
(see the appendix Cleaning, Disinfecting, Sterilizing).
! To avoid the risk of explosion, do not use flammable
anesthetic agents such as ether and cyclopropane in
the presence of the Scio module.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Gamma Series and is the answer not in the manual?

Dräger Infinity Gamma Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Gamma Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals