Default Settings and Biomedical Support
VF4 Infinity Gamma Series Page C-7
PLS Alarm Limits
(upper/lower)
120/45 150/50 170/80 Saved Setup
Temperature
Alarm Off Off Off Saved Setup
Alarm Limits
(upper/lower)
39/34 °C
102.0/93.2
°F
39/34 °C
102.0/93.2
°F
39/34 °C
102.0/93.2
°F
Saved Setup
Alarm Recording Off Off Off Previous
Patient
Units °C Previous
Patient
NBP
Alarm Off Off Off Saved Setup
Alarm Limits
(upper/lower)
Systolic
Mean
Diastolic
160/90
125/60
•110/50
160/90
125/60
•110/50
•80/50
•70/40
•60/25
Saved Setup
Alarm Recording Off Off Off Previous
Patient
Interval Mode Off Off Off Saved Setup
Calibration Mode Off Off Off Previous
Patient
Inflation Mode Adult:270 Ped.: 180 Neo.: 90 Previous
Patient
Measurement
Tone
Off Off Off Saved Setup
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient
discharge,
settings
return to...?