Page 2-8 Infinity Gamma Series VF4
STEPS: Charging Lead Acid Batteries with a
Battery Charger
You can charge up to four additional lead acid batteries with the
battery charger available from Dräger. To do so:
1. Place the lead acid batteries on the
battery charger.
2. Keep the lead acid batteries in the
charger for at least four hours to
ensure a full charge.
WARNING: Use the battery charger to charge lead acid
batteries only. Lithium ion batteries are not compatible
with the Dräger lead acid battery charger.
STEPS: Removing a Lead Acid Battery from the
Monitor
1. Connect the monitor to AC power or turn the monitor off.
NOTE: If the monitor is connected to AC power, monitoring can
continue while you replace the battery. If the monitor is not con-
nected to AC power, turn the monitor off before removing the bat-
tery in order to assure a proper monitor shut-down.
2. Place the monitor on a flat surface.
3. Open the battery compartment door (back of the monitor).
4. Lift the battery from under the locking tab and pull the bat-
tery out.
WARNIN G: To avoid explosion, do not disassemble, or
dispose of the battery in fire. Do not change the battery
in the presence of explosive hazards because of the pos-
sibility of sparks. If the monitor is not to be used for a
prolonged time, remove the battery from the battery
compartment.

Table of Contents

Related product manuals