Page 3-10 Infinity Gamma Series VF4
PICK AND GO versus Network Error
The network can distinguish between a network error and the
intentional removal of a bedside monitor from its Docking Sta-
tion.
If the monitor loses its connection with the network due to a tech-
nical problem, the network generates a network error message.
When the monitor is removed from its Docking Station for trans-
port, the central station displays a network status message (Bed
Disconnected).
If the bedside monitors speaker volume had been set to OFF dur-
ing network operation, upon loss of communication with the net-
work, it is automatically reset to 50% (default stand-alone level),
or to 100% in case of an active alarm (network error message).