Safety-related information
2.5.1 Calibration
For proper operation, never adjust the span before completing zero adjustment.
Performing these operations out of order will cause the calibration to be faulty.
If the intended operation is at high altitudes, the reading will be lower than the
reading at sea level (reduced partial pressure). A new span calibration is
recommended if the altitude or the ambient pressure is changed. The factory
calibration is set to sea level.
Dräger recommends calibrating instruments with target gas. This method is
more accurate than a surrogate gas calibration. A surrogate gas calibration may
only be performed as an alternative if a target gas calibration is not possible.
Methane and hydrogen should be calibrated with target gas only and not
calibrated with a surrogate gas.
For Polytron
®
8310 IR:
The user must also read the Instructions for Use for DrägerSensor IR, part
number 9023981
2.6 During operation
After modifying parameters with the PolySoft PC software or another software,
verify all parameters by downloading from the instrument or by checking them
on the Polytron
®
8xx0.
IP ratings do not imply that the equipment detects gas during and after exposure
to those conditions. In case of dust deposits and submersion/jet water, check the
calibration and the proper functioning of the device.
Risk of ignition of flammable or explosive atmospheres!
CAUTION: High off scale readings may indicate an explosive concentration.
(Only for instruments measuring combustible atmospheres.)
WARNING: Do not open when energized
Polytron
®
8100 EC-specific:
Response time t20 can exceed the limits of EN 50104 for small concentration
changes.
2.6.1 Maintenance
The maintenance intervals must be established for each individual installation.
Depending on safety considerations and application specific conditions the
instrument is used in, these might need to be shortened.
Refer to the maintenance section in the transmitter instructions for use.