EasyManuals Logo

Dräger Polytron 8000 Series Technical Manual

Dräger Polytron 8000 Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
28 Technical Manual | Dräger Polytron
®
8000 Series
Installation
1. Ground the enclosure of the instrument locally at the grounding lug, see the
following figure.
2. Connect the shield of the wires only to the instrument earth ground of the
controller (e g. chassis, ground busbar, etc.).
Unless special measures are taken (e. g. capacitive earthing), the shield must
only be connected at one end.
Fig. 3 Grounding the Ex d enclosure and the PCB unit
5.2.9 Closing the instrument
1. Ensure the following connections are properly made:
a. Wiring screws are tightened to the correct torque.
b. All cable connectors are secured with screws.
c. The sensor connector is plugged.
d. The grounding cable coming from the enclosure is connected to the lug on
the PCB unit, Grounding the Ex d enclosure and the PCB unit
2. Place PCB unit back into the enclosure.
3. Screw the lid back on, until it is seated with correct torque, and tighten set-
screw.
Fig. 4 Closing the Ex d enclosure with correct torques
5.3 Installing EC sensor
Observe the following figure.
This point is only valid for Polytron
®
8100 EC
1. If the instrument is already in operation, activate the function to change the
sensor (Sensor change function). Otherwise, a fault is displayed when the
sensor is deplugged.
38743
Without relay:
Polytron 8xx0
PWR+
PWR-
PWR-
4-20 mA
PE
40515
≥ 44 LB IN
≥ 5 Nm
≥ 8 LB IN
≥ 0.7 Nm
E

Table of Contents

Other manuals for Dräger Polytron 8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 8000 Series and is the answer not in the manual?

Dräger Polytron 8000 Series Specifications

General IconGeneral
Ingress ProtectionIP66/67
Detection PrincipleElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, toxic gases, and oxygen
DisplayLCD with backlight
Humidity Range0 to 99% RH (non-condensing)
Housing MaterialStainless steel
Output Signal4-20 mA, HART, RS-485 Modbus
Power Supply10 to 30 V DC
CertificationsATEX, IECEx, CSA
Temperature Range-40 °C to +65 °C (depending on sensor)

Related product manuals