EasyManuals Logo

Dräger Polytron 8000 Series Technical Manual

Dräger Polytron 8000 Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Technical Manual | Dräger Polytron
®
8000 Series 47
Calibration
9 Calibration
Obey the safety statements mentioned in 2.5 Commissioning!
A calibration checks and adjusts the measurement accuracy with a known test gas
concentration. First the zero point of the sensor and then the sensor span is
calibrated. Calibrations have to be performed on a regular basis. The length of
calibration intervals depends on the ambient conditions in which the sensor is
operated.
Ambient conditions and aging cause sensor drifts. Sensor drifts have a negative
impact on the measurement accuracy. Calibrations restore this accuracy.
Depending on the severity of the drift shorter calibration intervals have to be
established.
To evaluate the ambient conditions of a new installations, shorter calibration
intervals with documentation of the drift should be established. The plant operator
should establish customized calibration intervals with the obtained data.
Under normal conditions, Dräger recommends the following calibration intervals
2)
:
Electrochemical sensors (EC): 6-12 months
3)
Catalytic sensors (CAT): 4 months
Infrared sensors (IR): 6-24 months
3)
9.1 Test gases
For test gas properties (e. g. humidity, concentration) refer to the corresponding
sensor data sheet.
Test gas humidity is irrelevant for O
2 sensors.
Depending on the type of calibration different test gases are used.
Zero gas
Zero gas is a test gas to calibrate the zero point. If ambient air is free from
interfering impurities and measured gas, it can be used as zero gas. For O
2 and
CO
2 sensors, nitrogen (N2) is used.
Calibration gas
Calibration gas is a test gas to calibrate the sensor sensitivity during span
calibration. Calibration gas is a known concentration of the measured gas diluted
with clean air or nitrogen. For O
2 sensors no calibration gas is needed, as oxygen
from ambient air is used.
2) Observe sensor data sheet and sensor IfU
3) For applications in line with EN45544-1 the calibration interval must not exceed 6 months.

Table of Contents

Other manuals for Dräger Polytron 8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 8000 Series and is the answer not in the manual?

Dräger Polytron 8000 Series Specifications

General IconGeneral
Ingress ProtectionIP66/67
Detection PrincipleElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, toxic gases, and oxygen
DisplayLCD with backlight
Humidity Range0 to 99% RH (non-condensing)
Housing MaterialStainless steel
Output Signal4-20 mA, HART, RS-485 Modbus
Power Supply10 to 30 V DC
CertificationsATEX, IECEx, CSA
Temperature Range-40 °C to +65 °C (depending on sensor)

Related product manuals