Instrument settings
12.6 Changing the display contrast
1. Select Settings > Instrument > Display > Display contrast and confirm.
2. Change the contrast and confirm.
12.7 Changing the display mode
1. Select Settings > Instrument > Display > Display mode and confirm.
2. Select the desired mode and confirm.
If an alarm is triggered, the display shows the current gas concentration and the red
LED blinks regardless of the selected display mode.
12.8 Data logger settings
12.8.1 Information regarding the data-logger
The data logger can store up to 35000 values. At a sampling interval of 1
measurement per minute the data logger stores the measurement history of
approximately 24 days. This time can be significantly increased if the trigger
function is enabled.
The data logger can only be downloaded using the PolySoft PC software.
12.8.2 Displaying graph
The last 15 minutes as a time/concentration graph can be viewed.
1. Select Information > Data logger > Graph
12.8.3 Switching data logger on or off
1. Select Settings > Data logger > Datalogr. on/off and confirm.
2. Select Enable or Disable and confirm.
12.8.4 Setting the sampling time
This function defines how often a value is stored.
1. Select Settings > Data logger > Datalogr. on/off > Sample time and confirm.
2. Select the sample time and confirm.
12.8.5 Setting peak/average
This function determines which value is stored.
1. Select Settings > Data logger > Data logger > Peak/Average and confirm.
2. Select Peak or Average and confirm.
Standard Shows the standard display.
Non-display Shows the start screen and the applica-
ble symbols.