11 Maintenance
The maintenance intervals must be established for each individual installation.
Depending on safety considerations and application specific conditions the
instrument is used in, these might need to be shortened.
Every 6 months
Inspection by trained service personnel.
Check signal transmission to the central controller, LEDs and triggering of alarm
devices.
Calibrations have to be performed on a regular basis, see 9 Calibration
11.1 Performing a bump test
A bump test checks alarm activation without setting off the alarms.
1. Select Settings > Instrument > Bump test and confirm.
The maintenance symbol is displayed
The analog output is set to the maintenance signal.
2. Apply a known concentration of gas.
3. Press OK to return to measurement.
11.1.1 Checking response time (t90)
1. Perform a bump test and check the response time.
2. Compare the response time with the t90 values that are indicated in the
corresponding sensor data sheet.
CAUTION
Delayed response time at the gas detection control unit.
If the transmitter is connected to control units, the total response time could be
delayed. The entire measuring path must be taken into account (e. g. latency of the
control unit).
Ensure that the required response time is kept.
11.2 Sensor replacement
WARNING
Faulty Calibration
Faulty Calibration may cause inaccurate readings.
If the sensor is replaced, all settings and parameters must be checked for
correctness.
Check calibration to verify proper operation.
11.2.1 Polytron
®
8100 EC
A sensor can be replaced in the hazardous, classified area, without interrupting the
power supply.

Table of Contents

Other manuals for Dräger Polytron 8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 8000 Series and is the answer not in the manual?

Dräger Polytron 8000 Series Specifications

General IconGeneral
Ingress ProtectionIP66/67
Detection PrincipleElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, toxic gases, and oxygen
DisplayLCD with backlight
Humidity Range0 to 99% RH (non-condensing)
Housing MaterialStainless steel
Output Signal4-20 mA, HART, RS-485 Modbus
Power Supply10 to 30 V DC
CertificationsATEX, IECEx, CSA
Temperature Range-40 °C to +65 °C (depending on sensor)