Performance Approvals
Temperature -40/+65°C (-40/149°F)
Storage temperature 0/+40°C (32/104°F)
Enclosure protection
2)
IP 65/66/67
Range 0-100ppm
Accuracy (whichever is
greater)
measurement uncertainly (of meas. value) ±3 %
or minimum ±0.5ppm
Accessories
Dust filter 6812223
Adapter kit 6810536
Standard ANSI/ISA-92.00.01FM6340
1) HART communication shall not be used for safety-relevant communication
2) After submersion and jet water the sensor shall be replaced
DrägerSensor H2S LC (6809610)
Examination Type P8100 ETR 0400, 0401, 0500, 0501
Interface
1)
4-20mA, relay
Environmental parameters
Humidity 5 to 95%RH, non-condensing
Temperature -20/+65°C (-4/149°F)
Storage temperature 0/+40°C (32/104°F)
Enclosure protection
2)
IP 65/66/67
Range 0-100ppm
Accuracy (whichever is
greater)
measurement uncertainly (of meas.value) ±3 %
or minimum ±0.5ppm in the range >50ppm, ±10
% or ±3ppm
Accessories
Dust filter 6812223
Adapter kit 6810536
Standard ANSI/ISA-92.00.01FM6340
1) HART communication shall not be used for safety-relevant communication
2) After submersion and jet water the sensor shall be replaced
DrägerSensor H2S (6810435)

Table of Contents

Other manuals for Dräger Polytron 8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 8000 Series and is the answer not in the manual?

Dräger Polytron 8000 Series Specifications

General IconGeneral
Ingress ProtectionIP66/67
Detection PrincipleElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, toxic gases, and oxygen
DisplayLCD with backlight
Humidity Range0 to 99% RH (non-condensing)
Housing MaterialStainless steel
Output Signal4-20 mA, HART, RS-485 Modbus
Power Supply10 to 30 V DC
CertificationsATEX, IECEx, CSA
Temperature Range-40 °C to +65 °C (depending on sensor)