EasyManuals Logo

Dräger Polytron 8000 Series Technical Manual

Dräger Polytron 8000 Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Technical Manual | Dräger Polytron
®
8000 Series 53
Calibration
2. Select Calibration > Auto calibration and confirm.
The message please wait... is displayed and the instrument automatically
performs the zero calibration.
In case of an O
2 sensor Fresh air cal. is displayed.
The Maintenance signal is transmitted by the analog interface, no alarm or
fault relays are switched and the symbol is displayed.
3. After a successful zero calibration, the span calibration is started.
The parameters for the gas are displayed, e. g.:
6)
Gas : H2S
Unit: ppm
Concentr.:25
4. If the displayed parameters do not match the target gas, adjust the
parameters:
7)
a. Select Gas and confirm.
b. Select the calibration gas from the list and confirm.
c. Select Unit and confirm.
d. Select the unit of measurement from the list and confirm.
e. Select Conc. and confirm.
f. Set the concentration of the calibration gas.
5. If settings are correct:
a. Select Next and confirm.
A message like Gas flow ON H
2S is displayed.
6. Apply calibration gas.
a. Set gas flow corresponding to the sensor used (see "Gas flow for
calibrations", page 49)
7. Select Next and confirm.
The current value is displayed
After 15 minutes without confirmation, the instrument reverts to the calibration
menu without performing a calibration.
8. Once the current value is stable, the instrument automatically performs the span
calibration.
9. After a successful calibration, the new current value and the message Value
OK? are displayed.
If not, select Redo and confirm to perform the calibration again.
If the value corresponds the target value, proceed.
10.Turn off gas flow and remove the calibration adapter from the sensor or
disconnect tubing.
11. Select Accept value and confirm to finish the calibration. The instrument returns
to the calibration menu.
6) Not applicable for all sensors (see "Display, analog interface and relay status", page 41).
7) Only possible for certain sensors.

Table of Contents

Other manuals for Dräger Polytron 8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 8000 Series and is the answer not in the manual?

Dräger Polytron 8000 Series Specifications

General IconGeneral
Ingress ProtectionIP66/67
Detection PrincipleElectrochemical, Infrared, Catalytic Bead
Gases DetectedCombustible gases, toxic gases, and oxygen
DisplayLCD with backlight
Humidity Range0 to 99% RH (non-condensing)
Housing MaterialStainless steel
Output Signal4-20 mA, HART, RS-485 Modbus
Power Supply10 to 30 V DC
CertificationsATEX, IECEx, CSA
Temperature Range-40 °C to +65 °C (depending on sensor)

Related product manuals