MGR10 A/B/C - Revision : I (for 7.0 software revision) 35
Press on [OK] to go back to the previous page
Once the logging mode has been switched ON the measurement display will be in HOLD mode and
the bottom line of information text will show the total number of samples to be stored and the
number of measurements already stored.
Samples = 40/40
To start the logging press the [MES] key the measurements will start and the values stored. In
the CONTINUOUS trigger mode ,once the set number of values has been stored in the log the
measurements will be stopped.
In the Single trigger mode , press on the key [MES] for each measurement , after reaching the
preset number of readings , the next trigger will display the following message:
« Clear the current log ? »
Press [No] to keep the measurements and press on [LOG ] key to display the values .
Or press [Yes ] to start a new log session in clearing the previous measurements .