MGR10 A/B/C - Revision : I (for 7.0 software revision) 8
1.2 Definitions and terminology
Standard features of the MGR10 Microhmmeter include:
Direct display of resistance
Illuminated display
Up to 10 Amps measuring current (programmable)
Temperature compensation
Display hold function
Display zero function
4000 event data logger
Statistical reporting
Measuring current reversal
Hi/Lo Limits with Red/Green front panel warning LEDs
Digital filter for smoothing display
1.3 Principles of measurement
The MGR10 generates a constant current which is connected using the +I and -I leads across the
unknown resistance to be measured. This current also passes through a stable and accurate internal
standard. The volt drop across the unknown resistance is then measured together with the volt drop
across the internal standard, the ratio of these two measurements is then computed and displayed on
the front panel.