Serinus 40 User Manual 2.2  
 
 
3.  If the calibration fails then the instrument may have a hardware fault. The cell pressure PCA and 
the manifold pressure PCA have test points. To determine if the pressure sensor is faulty simply 
measure the voltage on the test points shown in the photos. The voltage measured across the 
test point is proportional to the pressure measured by the sensor, so if the sensor is exposed to 
ambient pressure at sea level then the voltage will be around 4 volts but if the sensor is under 
vacuum then the voltage will be low (e.g. 0.5 volts). If the test point measures zero or negative 
voltage then the assembly is most likely faulty and will need to be replaced. 
 
Figure 59 – Test point location 
 
Figure 60 – Typical test point reading of cell pressure with vacuum connected to exhaust port 
 
Figure 61 – Typical test point reading of manifold pressure sensor