Serinus 40 User Manual 2.2  
 
 
3.5.17  Valve Menu 
The Valve Menu allows the user to observe the opening and closing of valves as well as providing the 
ability to open and close them manually. 
When Enabled the instruments valves will open and close under 
microprocessor control. When Disabled the valves will change only 
in response to manual controls. 
Manually changing a valve while sequencing is enabled does not 
prevent the microprocessor from changing it again. 
Valve sequencing will remain off unless the instrument has 
returned to main screen for longer than 2 minutes. 
Shows the current operation of the NO
x
 select valve.  This valve 
switches the NO
2
 to NO converter sample to flow either through 
to the cell, or to bypass the sample out of the exhaust port of the 
instrument.  
Open = sample via NO
2
 converter.  
Closed = bypass NO
2
 converter sample out of instrument. 
Shows the current operation of the NO select valve. This valve 
switches the sample either straight through to the cell, or 
bypasses the sample out the exhaust port of the instrument. 
Selecting Switch will change the valve to the alternative state 
(Open = sample via sample port only, Closed = sample port via 
NO
2
 converter). 
Shows the current operation of the background select valve.  This 
valve determines whether the sample is pre-reacted with ozone 
before arriving at the cell. This creates no reaction in the cell and 
enables a cell ‘Background’ measurement to be taken. Selecting 
Switch will change the valve to the alternative state (Closed = 
sample as per normal, Open = react sample before cell).  
Shows the action of the valve that determines whether sample gas 
or calibration gas/internal zero air is being sampled (Open = 
Span/Zero, Closed = Sample Gas).  
Shows the action of the valve that determines whether the 
instrument calibration gas is taken from the Calibration Port or the 
Internal Zero (Closed = Internal Zero, Open = Calibration Port). 
Pressurised Span (optional) 
Shows the action of the valve that determines whether the 
instrument calibration gas is taken from the optional Pressurised 
Span Port (refer to Section 5.7). 
Pressurised Zero (optional) 
Shows the action of the valve that determines whether the 
instrument calibration gas is taken from the optional Pressurised 
Zero Port (refer to Section 5.7).