Serinus 40 User Manual 2.2
3.5.22 Data Logging Menu
This allows up to 12 parameters to be logged. After each
parameter (labelled Logging Param. 1 – Logging Param. 12),
place the number of the parameter that is to be logged. A 255
indicates the end of the list of parameters to be logged (refer
to Appendix A for a list of parameters).
This is an alternate and easier way to select logged
parameters. Instead of entering a number, select the item by
name from a list. Select the blank line to indicate the end of
the list of parameters to be logged.
Set the frequency that measurement data will be logged (1
second to 24 hours) or Disabled which means that no data is
logged to the USB memory stick.
Note: It takes about 1/3 of a second to log a measurement,
selecting the 1 second interval may slow down
communications when collecting logged data.
Note: The ‘–‘ key will delete the currently highlighted parameter; the ‘+’ key will insert a new
parameter at that location, moving the others down. The list of parameters must be
contiguous. Thus, when you delete a logging parameter, any parameters below it will be
moved up.
3.5.23 Serial Communications
This is the ID of the analyser when multi-drop RS232
communications is used. This defaults to 40 but can be
changed if multiple Serinus instruments are on the same
RS232 cable.
This is the ID of the NO
x
/NO
2
output (only available for
Bavarian protocol).
If Bavarian protocol is selected, then selection with NO
2
or
NO
x
should be output as the second gas.