6.6.3 Updating Firmware
It is important for optimal performance of the Serinus analyser that the latest firmware is loaded
onto the analyser. The latest firmware can be obtained by visiting Ecotech’s website:
http://www.ecotech.com/downloads/firmware
Or by emailing Ecotech at service@ecotech.com.au or intsupport@ecotech.com
To update the firmware from a USB memory stick, use the following procedure:
USB Memory Stick Update
1. Turn instrument off.
2. Place USB memory stick with new firmware (ensure that firmware is placed in a folder called
FIRMWARE) in the front panel USB Port.
3. Enter the Bootloader (Refer to Section 6.6).
4. Select option 3, (upgrade from USB memory stick) press 3 on keypad.
5. Wait till upgrade has completed.
6. Press 9 to start the analyser with new firmware.
6.6.4 Erase All Settings
This command is only required if the instrument’s firmware has become unstable due to corrupted
settings. To execute this command enter into Bootloader Menu and select key 4.
6.6.5 Start Analyser
The start analyser command will simply initiate a firmware load by pressing key 9 from the
Bootloader Menu. It is generally used after a firmware upgrade.