Serinus 40 User Manual 2.2
Page 52
Load Configuration
Loads a configuration file from the USB memory stick. Thus, you
can save a configuration and restore it later.
Auto-Backup
Selects whether the parameter and configuration files are
automatically saved once a day (at midnight).
Load Auto-Backup
Configuration
Loads the auto-backup configuration file. This is useful when the
configuration has been changed in error.
Instrument
This field allows the instrument to be set to either Online (normal
instrument operation) or In Maintenance (data is not valid, as
service work etc. is being performed).
Next Service Due
User editable field to set the date when the next instrument
service is required. Section 6 has a recommended maintenance
schedule that may be used as a basis for the interval entered
above. This value is also displayed as a non-editable field in the
Quick Menu (refer to Section 3.5.1).
Safely Remove USB Stick
This feature must be activated to safely remove the USB stick (also
found in the Quick Menu (refer to Section 3.5.1).
System Restart
Activating this will reboot the microprocessor.
3.5.15 Diagnostics
Digital Pots
Refer to Section 3.5.16
Valve Menu
Refer to Section 3.5.17
Tests
Refer to Section 3.5.18
Pressure/Temperature/Flow
Comp
Set to either On or Off:
On is used to compensate analyser measurements for
environmental fluctuations that might affect readings (pressure,
temperature and flow)
Off is used only when running diagnostics to see fluctuations in
readings
Diagnostics Mode
The instrument can be placed in 4 diagnostic modes:
Operate which leaves the instrument in normal operation mode
Optic which configures the instrument for tests on the optical
measurement source
Electrical which configures the instrument for testing of the
electrical circuits
Preamp which configures the instrument for testing of the pre-
amplification circuitry
Note: Selecting any mode other than Operate will disrupt
normal measurement.