Serinus 40 User Manual 2.2
Detects whether a USB memory stick is plugged into the front USB
port.
Indicates if the voltage of the concentration during background is
within the limits of the analog to digital converter (-0.26 to 3.29 V).
Indicates if the voltage of the concentration during NO
measurement is within the limits of the analog to digital converter
(-0.26 to 3.29 V).
Indicates if the voltage of the concentration during NOx
measurement is within the limits of the analog to digital converter
(-0.26 to 3.29 V).
3.5.5 Temperatures
Editable field to allow the user to change the current temperature
units of the analyser (Celsius, Fahrenheit, or Kelvin).
Editable field that sets the target temperature that some heated
components are regulated to.
Displays current temperature of the reaction cell.
Temperature of the NO
2
to NO converter.
Displays the air temperature inside the chassis, measured on the
main controller board.
Temperature of the cooled PMT block.
Displays the current temperature of the valve manifold.
3.5.6 Pressures & Flow
Select the units that the pressure will be displayed in (torr, PSI,
mBar, ATM, kPa).
Current ambient pressure.
Current pressure within the reaction cell.
Current pressure in the valve manifold.
The desired flow if the optional internal pump is installed.
Indicates the gas flow through the sample port of the instrument,
should be around 0.60.
Note: The sample flow will read 0.00 if there is a sample flow error.