Serinus 40 User Manual 2.2
Page 60
Background
Span
Zero
System
Assign the digital outputs 0-7 (refer to Figure 11 for pinouts).
3.5.28 Network Adaptor Menu
The Network Adaptor Menu allows the user to view or set the I.P. address, Netmask and Gateway if
the optional network port is installed.
To read the IP address, perform the following steps:
1. Set the instrument to Read IP.
2. Manually cycle power off.
3. Wait 3 seconds.
4. Turn power on.
5. Read or set the IP address.
Protocol (Network)
Sets the protocol used for the network port (EC9800,
Bavarian, Advanced, or Modbus). This must be set to
Advanced for Ecotech supplied software.
Start-up Mode
The following modes are available:
Normal: In this mode nothing is done with the network port
during boot-up. It is assumed to be configured correctly or
unused.
Read IP: This mode interrogates the network port for its IP
address. The menu will display the network address after
boot-up.
Set IP: You may enter an IP address, Netmask and Gateway
address (following the usual rules for formatting these
addresses). Please note that at this time the Serinus does not
validate the correctness of these entries.
When you cycle power, the Serinus will first instruct the
network port on its new address. It will then switch to Read IP
mode and read back the address it just set so that you may
verify it in the menu.
Set DHCP: This sets the network port into DHCP mode,
allowing the network to assign the Serinus an IP address.
IP Address
This is the current IP address of the analyser e.g. 192.168.1.2.