Serinus 40 User Manual 2.2
Page 72
Figure 22 New station connection
Table 6 Setting up a new station via Airodis
Property
Description
Station Name
The name of the station. If you have other loggers, the name is used to distinguish them.
Logger
Set this to “Ecotech Serinus” when downloading from any Serinus series instrument. This
will communicate with the Serinus on the Advanced protocol. If using a network or serial
connection, ensure that the Advanced protocol has also been selected on the Serinus itself
(Refer to Appendix A).
Time Zone
Set this to the time zone that the Serinus is configured to log in.
DST
Enable this option if you plan on changing the clock on the Serinus with daylight savings.
Leave this disabled if the clock does not shift during DST. Note: the Serinus will need to be
adjusted manually for DST it will not happen automatically.
Database Name
This is the name to be used for the table in the SQL database containing this station’s data.
It must be unique for each station.
Device ID
Enter the multidrop ID of the Serinus (Refer to Section 4.1). If you are not using multidrop,
this can be set to “0” or left blank.
Link Type
Select the type of connection used to connect to the Serinus. Additional properties will
appear to connect to the Serinus. Make these align with the configured communications
settings on the Serinus.
Log Interval
This needs to be the same as the Data Log Interval setting on the Serinus.
The available fields for communications parameters will change when you change the link type. You
will need to set the communication parameters that have been defined on the Serinus.
5. Once the station has been created, save the station by clicking the Save shortcut icon or
File→Save.
6. Click Acquire Configuration. This will probe the Serinus for a channel list. After a few seconds,
the channel list should be visible in the Channels tab.