9. Push the analyser fully in. At this point, make sure that the analyser slides reach and locate in to
the plastic catches at the rear end of the outer slides. Adjust the outer slides as required until this
is achieved. Remove analyser and now tighten the M4 screws and nuts that secure the front and
rear claws on both sides of the rack.
To Remove the Instrument
1. To remove the instrument first pull instrument forward of rack giving access to the slides.
2. Find the rack slide lock labelled “Push” and push it in whilst sliding the instrument out of the rack,
complete this for both sides whilst carefully removing instrument.
Figure 80 Rack installation
8.3 High Pressure Zero/Span Valves
High pressure span calibration valve (factory installed) PN E020108
High pressure zero calibration valve (factory installed) PN E020109
Note: This is not intended as a source for calibrating the instrument. This should only be used as an
operational check of the instrument’s zero point and single upscale point (recommended as 80% of
full scale).
Please refer to Section 5.7 for operation of this installed option.