Installation
Page 37
2.5 EN Type Approval Set-up
The Serinus 40 has been certified to MCERTS performance standards for Continuous Ambient Air
Quality Monitoring Systems. The certificate number is Sira MC 100167/02. The Serinus 40 must be
used under the following conditions to meet EN requirements:
Range (NO)
0-1000 ppb
Range (NO
2
)
0-260 ppb
Ambient Temperature
0-30°C
The analyser must be operated and maintained in accordance with this user manual.
The following menu selections must be used:
Calibration Menu
Span comp: Disabled
ServiceDiagnostics Menu
Press/Temp/Flow comp: On
Diagnostic mode: Operate
Control loop: Enabled
2.6 Transporting/Storage
Transporting the Serinus should be done with great care. It is recommended that the packaging the
Serinus was delivered in should be used when transporting or storing the instrument.
When transporting or storing the instrument the following points should be followed:
1. Turn off the instrument and allow it to cool down.
2. Remove all pneumatic, power and communication connections.
3. If storing over a long period (6 months) turn the battery off by switching the switch on the main
processor board (refer to Figure 5) to the left.
4. Remove the instrument from the rack.
5. Replace the red plugs into the pneumatic connections.
6. Place the instrument back into a plastic bag with desiccant packs and then seal the bag (ideally
the bag supplied upon delivery).