Serinus 40 User Manual 2.2
<SP><ETC><bcc1><bcc2>
Where:
<SP> = space (0x20 hex)
kkk = device’s multi-drop ID. If the DA command is issued without an ID, then the response omits this
field. Exception: the S40 family always includes both ID fields, even when a DA command without an
ID is issued.
+nnnn+ee = main instantaneous gas concentration (for S40 family, this is NO)
ss = status byte with the following bit map:
Table 16 – Bit map
Instrument off (this value is always set to 0)
Units: 1 = Volumetric, 0 = Gravimetric
Background mode (S30 and S50 family only)
ff = failure byte for both channels with the following bit map (positive logic):
Table 17 – Bit map (positive logic)
Lamp failure (S40 family only)
Cell heater failure (S30, S40 and S50 family only)