1.  Introduction 
1.1  Description 
The Serinus 40 Oxides of Nitrogen Analyser uses gas phase chemiluminescence detection to perform 
continuous analysis of nitric oxide (NO), total oxides of nitrogen (NOx) and nitrogen dioxide (NO
2
).  
The Serinus 40 analyser measures to a sensitivity of 0.4 ppb with a range of 0-20 ppm. The 
measurement of these three gases is performed with the following main sub-components: 
  NO
2
 to NO converter 
  Delay coil (NO
X
 channel) 
  Ozone generator 
  Reaction cell 
  Photomultiplier tube (PMT) 
A microprocessor programmed with Serinus firmware monitors the detector response and many 
other parameters. The concentrations are automatically corrected for gas temperature and pressure 
changes and when using gravimetric units (e.g. µg/m³) is referenced to 0°C, 20°C or 25°C at 1 
atmosphere. This allows the Serinus 40 to provide readings in the relevant units for your 
requirements. 
The U.S. EPA has designated the Serinus 40 Oxides of Nitrogen Analyser as a reference method and 
SIRA has designated it as an EN approved instrument. 
1.2  Specifications 
1.2.1  Measurement 
Range 
0-20 ppm auto ranging.  
USEPA designated range: 0-0.5 ppm.  
MCERTS EN certification ranges: NO (0 to 1,000 ppb) NO
2
 (0 to 260 ppb). 
Lower detectable limit: 0.4 ppb, with Kalman filter active. 
1.2.2  Precision/Accuracy 
Precision 
0.4 ppb otherwise 0.5% of reading, whichever is greater. 
Linearity 
±1% of full scale.