Advanced Protocol Parameter List
Page 129
Appendix A. Advanced Protocol Parameter List
Note: Parameters are for all Serinus series analysers and may not be applicable to an individual
analyser.
Table 12 - Advanced protocol parameter list
#
Description
Notes
0
Sample / Cal Valve
0=sample, 1=cal/zero
1
Cal / Zero Valve
0=zero, 1=cal
2
Internal Span Valve
0=closed, 1=open
3
Spare Valve 1
0=closed, 1=open
4
Spare Valve 2
0=closed, 1=open
5
Spare Valve 3
0=closed, 1=open
6
Spare Valve 4
0=closed, 1=open
7
NO
x
Measure Valve
0=NO, 1=NO
x
8
NO
x
Bypass Valve
0=NO, 1=NO
x
9
NO
x
Background Valve
0=closed, 1=open
10
Valve Sequencing
0=off, 1=on
11
LCD Contrast Pot
0=lightest, 255=darkest
12
SO
2
Reference Zero Gain Pot
S50 reference zero pot
13
CO Measure Gain Pot
S30 measure gain adjust
14
CO Reference Gain Pot
15
CO Test Measure Pot
See 149. Exists
16
PMT High Voltage Pot
High voltage controller Pot for PMT S50 & S40
17
SO
2
Lamp Adj. Pot
S50 lamp adjust pot
18
O
3
Lamp Adj. Pot
S10 lamp adjust pot
19
O
3
Zero Measure Pot: Coarse
S10 signal zero (coarse)
20
O
3
Zero Measure Pot: Fine
S10 signal zero (fine)
21
PMT Fan Pot
PMT fan speed controller pot
22
Rear Fan Pot
Chassis fan speed controller pot
23
Pump Speed Motor Driver Pot: Fine
Internal pump speed fine pot
24
Pump Speed Motor Driver Pot: Coarse
Internal pump speed coarse pot
25
Analogue Input 0
SO
2
reference signal
26
Analogue Input 1
CO reference signal
27
Analogue Input 2
O
3
reference signal
28
Analogue Input 3
SO
2
& O
3
lamp current